Type: Article
Design and development of radioactive xenon gas purification and analysis system based on molecular sieves
Journal: Journal of Environmental Radioactivity (18791700)Year: October 2018Volume: 190Issue: Pages: 66 - 72
DOI:10.1016/j.jenvrad.2018.05.004Language: English
Abstract
The dynamic adsorption of xenon on molecular sieve packed columns was investigated. The modified Wheeler-Jonas equation was used to describe adsorption parameters such as adsorption capacity and adsorption rate coefficient. Different experimental conditions were accomplished to study their effects and to touch appropriate adsorbing circumstances. Respectable consistency was reached between experimental and modeled values. A purification and analysis setup was developed for radioactive xenon gas determination. Standard sample analysis results approved acceptable quantification accuracy. © 2018